OMS logo

OMS Blog

Equipment Calibration Management Best Practices for Testing Labs

Why overdue calibrations keep appearing in NATA and NABL assessments — and the management practices that prevent them.

·Jayant Chandavarkar

Overdue equipment calibration is, year after year, one of the most common findings in NATA and NABL assessments. In a well-run laboratory, it shouldn't happen — calibration due dates are known in advance, scheduling a calibration is not complicated, and the consequences of missing a date are significant. Yet it happens routinely, almost always for the same reason: the tracking system isn't reliable enough to catch it before the assessor does.

Here are the calibration management practices that prevent it.

1. Maintain a Complete Equipment Register

The foundation of calibration management is a complete, current list of every instrument that requires calibration — with its unique equipment ID, calibration interval, last calibration date, next due date, and the accredited calibration provider used. If this register isn't complete, you can't manage what you can't see.

Common gaps in equipment registers:

  • New instruments added to the lab but not yet added to the register
  • Reference standards included but working standards (balances, thermometers, gauges used daily) omitted
  • Instruments currently "out for calibration" removed from the active register rather than flagged as pending
  • Retired instruments not formally decommissioned — still listed as active but not being calibrated

Every instrument that influences a measurement result must be in the register. OMS Software's asset management module maintains this register centrally, with status tracking from active through to decommissioned.

2. Set Calibration Intervals Based on Risk, Not Habit

ISO/IEC 17025 Clause 6.4 requires that calibration intervals be established. What it doesn't prescribe is a specific interval — that's the laboratory's professional judgement. Many labs default to annual calibration for everything because it's simple. But a critical reference standard used multiple times daily warrants more frequent calibration than a thermometer checked monthly.

Factors to consider when setting intervals:

  • Frequency of use — instruments used more often drift more
  • Stability history — if historical calibration records show consistent results, a longer interval may be justified; if results vary significantly, shorten the interval
  • Consequence of out-of-tolerance — for instruments influencing safety-critical results, err toward shorter intervals
  • Manufacturer recommendations — a starting point, not a mandate

3. Automate Due Date Alerts — Don't Rely on Manual Checking

The most common reason for overdue calibrations is simple: nobody checked. The spreadsheet wasn't opened, the email reminder was buried, or the person responsible was on leave and nobody else picked it up. Manual checking of calibration due dates is a brittle system that fails under normal operational pressure.

Automated alerts — sent to the equipment owner and the quality manager at 30, 14, and 7 days before due — change this completely. The alert arrives without anyone having to initiate it. Missing a calibration requires actively ignoring multiple reminders, which is much less likely than forgetting to check a spreadsheet. This is exactly the kind of control that separates labs that stay compliant from labs that discover overdue instruments during an assessment.

4. Record What Happens When Equipment Goes Out of Tolerance

When a calibration reveals that an instrument was operating outside its acceptable tolerance — an out-of-tolerance (OOT) result — the laboratory is required to assess the impact on previous test results. This is one of the most serious calibration findings because it raises the question: how long has this instrument been out of tolerance, and which results were produced with it?

The required response is a formal non-conformance investigation:

  • Determine the date of the last known in-tolerance calibration
  • Identify all test results produced using this instrument since that date
  • Assess whether those results are affected and whether clients need to be notified
  • Document the investigation, findings, and any client notifications with objective evidence

Labs that have this process documented in advance handle OOT findings efficiently. Labs that don't tend to manage the response inconsistently — which creates further findings. See our guide on improving traceability in laboratory testing for how result-to-instrument linkage supports this investigation.

5. Ensure Calibration Certificates Meet ISO/IEC 17025 Requirements

Not all calibration certificates are equal. A compliant calibration certificate for a reference standard used in an accredited laboratory must include:

  • Measurement results with expanded uncertainty (not just a pass/fail statement)
  • Reference to the calibration method used
  • Evidence that the calibration was performed by a NATA-accredited (or equivalent) laboratory
  • Clear identification of the instrument calibrated (serial number, model)

Certificates from uncredited suppliers, or certificates that only state "pass" without measurement data and uncertainty, do not satisfy metrological traceability requirements under ISO/IEC 17025 Clause 6.4.6. This is one of the most frequently cited non-conformances in laboratory assessments.

The full value of calibration management isn't just keeping instruments current — it's being able to demonstrate, for any test result, that the instrument used was in calibration at the time. This linkage is the metrological traceability chain that ISO/IEC 17025 Clause 6.4 requires.

In a purpose-built LIMS, this linkage is automatic: the instrument ID is recorded with the test entry, and the calibration status is verifiable from the equipment register. In a spreadsheet-based system, it requires manual cross-referencing that is rarely done proactively — only under assessment pressure.

Manage Calibration Automatically with OMS

OMS tracks every instrument's calibration due date, sends automated alerts, manages out-of-tolerance investigations, and links calibration records to test results — all in one platform.

Book a Free Demo

Frequently Asked Questions

How often should laboratory equipment be calibrated?
ISO/IEC 17025 requires laboratories to establish calibration intervals but does not prescribe specific frequencies. Intervals should be based on risk: frequency of use, stability history from prior calibrations, and the consequence of an out-of-tolerance result. Many labs default to annual calibration, but high-use reference standards may warrant shorter intervals.
What is an out-of-tolerance calibration result and what must I do?
An out-of-tolerance (OOT) result means the instrument was operating outside its acceptable measurement limits. The laboratory must formally assess the impact on all test results produced with that instrument since its last known in-tolerance calibration, determine if clients need to be notified, and document the investigation with objective evidence as a non-conformance.
What must a compliant calibration certificate include under ISO/IEC 17025?
A compliant certificate must include: measurement results with expanded uncertainty (not just pass/fail), the calibration method used, evidence that the calibration was performed by a NATA-accredited or equivalent laboratory, and clear identification of the instrument calibrated by serial number and model. Certificates that only state 'pass' without measurement data do not satisfy Clause 6.4.6.
Why do calibration overdue findings keep appearing in NATA assessments?
Almost always because due dates are tracked manually — in spreadsheets or calendars that nobody actively monitors. The tracking system fails under normal operational pressure: the person responsible is on leave, the spreadsheet isn't checked, the email reminder is buried. Automated alerts triggered by the system at 30, 14, and 7 days before due date remove the reliance on human memory.
How does calibration management link to metrological traceability?
ISO/IEC 17025 requires that every test result be traceable to the instrument used, and that instrument be traceable to national measurement standards through an unbroken calibration chain. This means the calibration status of an instrument at the time of testing must be verifiable — which requires instrument ID to be recorded with each test result and the calibration record to be accessible from the same system.